The LTV 900 features many of today's most popular modes of ventilation, including pressure support, NPPV, adjustable flow triggering and an apnea back-up mode. Its versatile 13.4 lb (6.1 kg) design provides patients with the freedom to be more mobile. Features and Benefits Ease of UseAn intuitive interface makes the LTV 900 easy to understand and operate. The unit features extensive patient monitoring capabilities and a comprehensive alarms package for optimum patient care. With a footprint and size at a fraction of comparably equipped ventilators, the LTV 900 allows for greater patient mobility. Patient ComfortDesigned to provide optimum patient comfort and care, the LTV 900 features: Variable Rise Time Variable Sensitivity Flow Trigger Variable Flow Termination Variable Time Termination Leak Compensation NPPV Mode High Pressure Alarm Delay Oxygen Bleed InLow Flow / Low Pressure Versatile Power OptionsFor increased patient mobility and ease of transport, the LTV can be used with a variety of power sources: AC power LTV Transport Battery System LTV Universal Power Supply 3-Hour External Battery 9-Hour External Battery Automobile Lighter Adapter.

Basic Description  
 
Manufacturer:  Pulmonetic Systems
Ventilator Type:  Positive Pressure
Invasive
Noninvasive
Conventional
Ventilator Class:  Transport
Subacute
Home
CPAP
Patient Population Range:  Pediatric
Adult
Acute
Long-term (chronic)
 
Characteristics
 
Modes--Invasive:  Volume-controlled modes --- ASSIST/CONTROL; SIMV (PSV); CPAP
Modes--Non-Invasive:  CONTROL; ASSIST/CONTROL; SIMV; APNEA BACKUP
Operating Range of Controls:  Breath Rate: 0 to 80 BPM;
Tidal Volume: 50 to 2000 ml;
Inspiratory Time: 0.3 to 9.9 sec. (100 lpm);
PS/Spontaneous Flow: >/= 140 lpm;
Pressure Support: Off, 1 to 60 cmH2O;
Sensitivity: Off, 1 to 9 lpm;
Bias Flow: 10 lpm during exhalation;
PEEP/CPAP: 0 to 20 cmH2O
Graphic Display:  Real time waveform of Pressure, Flow and Tidal Volume with Freeze and cursors;
Real time loops of Pressure/Volume and Flow/Volume with Reference Loop and up to 4 historical loops;
Provides trending for 24 hours on f, Spf, Vte, VE, PIP, MAP, PEEP, Static Compliance and Auto PEEP;
Displays ventilator settings and monitored values;
Communicates with Patient Data Systems;
Available print on all screens;
Touch Screen color panel that can be custom configured
Size:  3" x 10" x 12" (8 x 23 x 30 cm)
Weight:  12.4 lbs (5.66 kg)
Other Characteristics:  PS Variable Rise Time: 0.1 to 0.9 sec.;
PS Variable Flow Termination: 10% to 40%;
PS Variable Time Termination: 1.0 to 3.0 sec.;
Remote Alarm Connection;
Has a cigarette lighter adapter
Characteristics (Non - Technical)
Approvals:  ETL/cETL Listed, IEC 601.1, ISO 9001/EN 46001 Registration, BSI Registered FM 45841
Other Comments:  20,000 hour service plan: 2-10,000 hour maintenance; 30,000 hour service plan: 2-10,000 hour and 1-30,000 hour maintenance
 
Specifications
Front Panel Controls:  Power: On/Standby;
Ventilation Modes: Control, Assist/Control, SIMV, CPAP;
Breath Types: Volume Control, Pressure Support, Spontaneous;
Breath Rate;
Tidal Volume;
Inspiratory Time;
PS/Spontaneous Flow;
Pressure Support;
Sensitivity;
Bias Flow;
PEEP/CPAP;
Control Lock: Easy or Hard unlock methods;
Manual Breath: 1 x current settings
Power Requirements:  Electrical Requirements:
11 to 15 VDC;
90 to 250 VAC;
47 to 63 Hz;
Pneumatic Requirements; 
Low Pressure Source: < 80 lpm
Internal Battery: 60 mins.
External Battery: 3 Hr or 9 Hr. Source Flow 
Oxygen (Medical Grade): 100 L/min 
Languages:  English, French, German, Spanish, Italian, Swedish, Danish, Japanese
Alarms:  High Pressure Audible
Alarm Delay: No Delay, 
1 Breath, 2 Breaths
Alarm Silence/Reset: 60 seconds; 
Ventilator Inoperative: Red LED; 
Disc/Sense; 
Low and Lost External Power; 
Low and Empty Internal Battery; 
High and Low Oxygen Inlet Pressure
Alarm Controls:  Apnea Interval: 10 to 60 seconds; 
High Pressure Limit: 5 to 100 cmH2O; 
Low Peak Pressure: Off, 1 to 60 cmH2O; 
Low Minute Volume: Off, 0.1 to 99 liters; 
Alarm Volume: 60 to 85 dBA at 1 meter
Non-Adjustable Alarms:  Disconnect;
Low & Lost External Power;
Low & Empty Internal Battery;
High & Low Oxygen Inlet Pressure;
Hardware Fault;
Transducer Fault;
Vent Reset;
Setting Defaults
AC Power and Battery Indicators:  Red, Green, Yellow LEDs
Other Specifications:  Oxygen Source: Low pressure bleed in;
Bias Flow: 10 lpm during exhalation;
Over Pressure Relief: 110 cmH2O;
Sub-Ambient Relief: Pressure Drop: -5 cmH2O;
Control Lock: Easy or Hard Unlock methods;
Manual Breath: single volume or pressure control breath
NPPV Mode: Alarm package limited to: High Pressure, Disconnect, Defaults, Power Lost, Batter Low, Battery Empty, Apnea, Ventilator Inoperative

Extended Features:
NPPV Mode;
Apnea Back-up Ventilation;
Pressure Support Variable Rise Time;
Pressure Support Variable Flow Termination Criteria;
Pressure Support Variable Time Termination;
Leak Compensation

Other Monitored Variables:
IE Ratio: 99:1 - 1:99;
Calculated Peak Flow: 10 - 100 lpm;
Patient Effort: Green LED;
Monitor Scrolling
Monitored Variables :
Pressure Peak Inspiratory Pressure: 0 - 120 cmH2O, Mean Airway Pressure: 0 - 99 cmH2O, PEEP: 0 - 99 cmH2O, Airway Pressure Display: -10 to 108 cmH2O
Respiratory Rate Total Breath Rate: 0 - 250 bpm
Tidal Volume Exhalation Tidal Volume: 0 - 4000 ml
Minute Volume Total Minute Volume: 0 - 99.9 liters